A specific flow meter which was calibrated for air, normal density 1.293 kg/m? is to be used for nitrogen, normal density 1.25 kg/m*. Read the factor 1.02 (heavy border) from the intersection of Air column and the Nitrogen row. The flow rate values indicated by the flow meter are to be multiplied by this factor.

Table 10 A specific flow meter which was calibrated for air, normal density 1.293 kg/m? is to be used for nitrogen, normal density 1.25 kg/m*. Read the factor 1.02 (heavy border) from the intersection of Air column and the Nitrogen row. The flow rate values indicated by the flow meter are to be multiplied by this factor.