First Experimental Results with the ClearPEM Detector
IEEE Nuclear Science Symposium Conference Record, 2005
https://doi.org/10.1109/NSSMIC.2005.1596639Abstract
First experimental results of the imaging system Clear-PEM for positron emission mammography, under development within the framework of the Crystal Clear Collaboration at CERN, are presented. The quality control procedures of crystal pixels, APD arrays and assembled detector modules are described. The detector module performance was characterized in detail. Results on measurements of light yield, energy resolution, depth-of-interaction and inter-channel cross-talk are discussed. The status of the development of the front-end electronics and of the data acquisition boards is reported. (PEM) prototype under development within the framework of the Crystal Clear Collaboration . The detector is based on pixelized LYSO:Ce crystals optically coupled on both extremities to avalanche photodiodes (APD) and readout by a fast low-noise electronic system. A dedicated digital trigger and data acquisition system is used for on-line selection of coincidence events with high efficiency, large bandwidth and negligible dead-time. The scanner consists of two compact and planar detector heads with adequate dimensions for breast Manuscript
References (8)
- C. J. Thompson et al., Positron Emission Mammography (PEM): a promising technique to detect breast cancer, IEEE Trans. Nucl. Sci. vol. 42, pp.1012-1017, 1995.
- W. W. Moses, Positron emission mammography imaging, Nucl. Instrum. Meth. vol. A 525, pp. 249-252, 2004.
- P. Lecoq and J. Varela, Clear-PEM, a dedicated PET camera for mam- mography, Nucl. Instrum. Meth. vol A 486, pp.1-6, 2002.
- M. C. Abreu, Design and evaluation of the Clear-PEM scanner for positron emission mammography, IEEE Trans. Nucl. Sci. (accepted).
- J. Trummer et al., Scintillation properties of LuYAP and LYSO crystals measured with MiniACCOS, an automatic crystal quality control systems, IEEE 2005 NSS-MIC Conference, Puerto Rico, USA.
- E. Albuquerque et al., The Clear-PEM electronics system, IEEE Trans. Nucl. Sci. (accepted).
- P. Rodrigues et al., Performance simulation studies of the Clear-PEM DAQ/Trigger system, in Conference Records of the 14th IEEE-NPSS RT2005, Stockholm, Sweden.
- C. Leong et al., Design and test issues of a FPGA based data acquisition system for medical imaging using PEM, in Conference Records of the 14th IEEE-NPSS RT2005, Stockholm, Sweden.